99
100
Steps
Do This
Decide whether to add
16.
another block, delete
the block or continue to
set BG targets.
If you want to
Then do this
Add another
block
Notice
1.
Add block is highlighted.
Press
2.
Enter button.
Repeat Steps 14 to 16 to add additional blocks.
3.
Delete the
block
Scroll to highlight
1.
Delete.
Press
2.
Enter button.
Scroll to
3.
Yes to confirm block deletion.
Press
4.
Enter button.
Continue to set
BG target
s
Note:
You must add at least 1 block before continuing.
Scroll to highlight Next if not already highlighted.
Press Enter button.
Steps
Do This
Set BG TARGETS.
17.
WARNING:
Accurate determination and programming of blood glucose targets is
necessary to avoid possible delivery errors that may result in adverse medical events
or serious injury.
Notice
a.
Time line and dashes are highlighted.
Press
b.
Enter button.
Notice
c.
Time is highlighted.
Press
d.
Enter button.
Scroll to the desired time.
e.
Press
f.
Enter button.
Notice
g.
BG Target is highlighted.
Press
h.
Enter button.
Scroll to the desired blood glucose target.
i.
Note:
The default setting is 110 mg/dL (6.1 mmol/L), and the range is
50 mg/dL to 300 mg/dL (2.8 mmol/L to 16.7 mmol/L).
Press
j.
Enter button.
Check that the time and
18.
BG Target settings are
correct.
Review the
a.
Time and BG Target settings on the screen.
Scroll up to select and change any setting as needed.
b.
Notice
c.
Next is highlighted.
Press
d.
Enter button
Section 6: When You’re Ready – Optional Settings
6.1 Set Up Additional Options – SETUP WIZARD
Section 6: When You’re Ready – Optional Settings
6.1 Set Up Additional Options – SETUP WIZARD
DOC15736_Rev-A-Man.indd 99-100
2/9/09 11:29:43 AM