13
14
Procedure:
Steps
Do This
Prepare to install a new
1.
battery into your pump
.
If you are
Then do this
Already wearing the
pump
Stop basal
1.
delivery. See Section 3.1.1 for
procedure.
Disconnect your pump and infusion set from
2.
your body. Follow the procedure described in
your infusion set’s instructions to disconnect the
infusion set.
Note:
You may only need to detach the cannula from
the cannula housing.
Go to Step 2.
3.
Setting up the pump
for the first time
Go to Step 2.
Obtain a fresh AAA alkaline
2.
battery.
Be ready to install a fresh battery within 15 minutes of removing the old
battery.
Warning:
A 15 minute delay may delete your time and date settings and
Insulin
on Board (IOB)
calculations. Your pump may also lose the history of any recent
boluses. Make sure the fresh battery is at room temperature to avoid false
alarms.
Remove the battery
3.
compartment cap.
Find the battery compartment on the side of your pump.
a.
With the pump screen facing you, as shown in the picture, place your
b.
thumb on the groove and turn the cap counter-clockwise until the cap
pops out.
If necessary, gently pull the battery cap off.
c.
Decide what to do next.
d.
If you are
Then do this
Removing an old battery
Go to Step 4.
Installing a fresh battery
Go to Step 5.
Steps
Do This
Remove the battery.
4.
If necessary, tap the pump against your hand or gently pull to remove the
battery.
The European Battery Directive requires separate collection of
spent batteries, aiming to facilitate recycling and to protect the
environment. The batteries in this product should be removed
and disposed in accordance with local regulations for separate
collection of spent batteries.
Insert a fresh battery.
5.
AAA
Insert the battery
negative end
(-) first.
Replace the battery
6.
component cap.
Carefully place the metal prongs of the battery cap into the space
a.
between the battery and the inside of the pump compartment at a 90
degree angle.
Push down and use your thumb to press the cap in place.
b.
Twist cap clockwise to close it securely.
c.
Caution:
Be careful
NOT
to use too much force to place the battery cap on the
pump. Excess force can lead to bending or breaking of the metal prongs.
Section 1: Getting Acquainted
1.4 Install Pump Battery
Section 1: Getting Acquainted
1.4 Install Pump Battery
DOC15736_Rev-A-Man.indd 13-14