CONFIGURATION
PEN POSITION
CONFIGURATION
POSITION PEN 1 TO
CHART ZERO
POSITION PEN 1 TO
CHART FULLSCALE
PEN CAL ENABLE
YES
NO
NO
3.4.3
Pen Position Configuration Page
Information.
•
Allows accurate calibration of the pen positions on the chart.
•
Can be used to remove effects of inconsistencies in the charts.
Pen Calibration Enable
Select
YES
to advance to the zero and full scale setting parameters.
Chart Zero Position
The chart advances at high speed and the pen moves to its zero position. With
the door open set the chart zero position of pen 1 – Magenta/Red using the
and
keys. Press the
key to store the pen 1 zero position.
Chart Full Scale Position
The chart continues to advance at high speed and the pen moves to its full
scale setting. Set the chart full scale position of pen 1.
Press the
key to store the pen 1 full scale position.
Return to
Pen Position Configuration
frame.