ProKiln GAC400
PROBE GAS SAMPLING SYSTEM | OI/GAC400-EN REV. A
33
Parameter input Alarm setpoint menu
Figure 29: Alarm setpoint menu
Parameter input Function
Default value
T01 TAH
Heated probe temperature high
maintenance request
200 °C
T01 TAL
Heated probe temperature low
maintenance request
170 °C
T02 TAH
Heated hose temperature high
maintenance request
190 °C
T02 TAL
Heated hose temperature low
maintenance request
150 °C
T03 TAH
Heated shut-off valve temperature
high maintenance request
190 °C
T03 TAL
Heated shut-off valve temperature low
maintenance request
150 °C
T10 TAHH
Water tank temperature high-high
alarm
80 °C
T10 TAL
Water tank temperature low
maintenance request
5 °C
T11 TAHH
Water temperature high-high forward
cooling water flow alarm
80 °C
T12 TAHH
Water temperature high-high return
cooling water flow alarm
90 °C
F12 FAL
Water flow return low maintenance
request
80 l/m
F12 FALL
Water flow return low-low alarm
60 l/m
T12-T11 TAHH
Temperature difference (forward-
return) high-high alarm
25 °C