ProKiln GAC400
PROBE GAS SAMPLING SYSTEM | OI/GAC400-EN REV. A
… 3 Design and function
… Default operation parameter settings
Parameter input M1 Motor Control menu
Figure 25: M1 Motor Control menu
Parameter input Function
Default value
Torque limit
Protective limit for motor and gear
65%
M1 Run
Motor run
Status of motor
M1 Switch On
Operation allowed
The motor can run
Stop Coast
Stop at retracted position
On
Stop Ramp
Run direct speed control
On
M1 Break release Release the brake to move buggy
Off
Horn On
The horn will sound during movement
of probe
On
Failure sec.
Maximum time allowed for the probe
to move
60 sec.
Parameter input - PID H1 Probe heater Setup menu
Figure 26: PID H1 Probe heater Setup menu
Parameter input Function
Default value
Setp.
Temperature set-point of sample tube
inside of the probe
180 °C
Reset
Reset to factory setting
On/Off
Power control for heater
On
Parameter input PID H2 Heated hose setup menu
Figure 27: PID H2 Heated hose setup menu
Parameter input Function
Default value
Setp.
Temperature set-point of heated hose
inside of the retractor
180 °C
Reset
Reset to factory settings
On/Off
Power control for heater
On
Parameter input PID H3 Heated valve setup menu
Figure 28: PID H3 Heated valve setup menu
Parameter input Function
Default value
Setp.
Temperature set-poin of heated shut-
off valve (if available)
180 °C
Reset
Reset to factory settings
On/Off
Power control for heater
On