Navigator 500
Sodium analyzer
7 Menu descriptions
46
OI/ASO550–EN Rev. G
7.1.2 Sensor Setup
Used to set the wet-section tag, measurement units, operational and stream ranges and filter type
and enable / disable the
Flow Meter
(optional) and
Manual Valve Control
functions.
Menu
Sensor Setup
Exit
Select
Menu
Comment
Default
Sodium 1 – 4
The identity of the sodium wet-section being set up (if a sodium
multi-steam wet-section,
Sodium
only is displayed).
Tag
Note
: Single-stream wet-sections only.
Sets the (colored) 16-character alphanumeric wet-section tag
displayed on
Operator Pages
.
TAG 1
Stream 1 (3) Tag
Note
: Multi-stream wet-sections only.
Sets the 16-character alphanumeric tag for each selected stream
displayed on
Operator Pages
.
Units
Selects the measurement units:
ppb / µg/l / µg/kg.
ppb
Range High
Note
. Single-stream wet-sections only.
Sets the range span in
Chart
and
Bargraph
views.
Stream 1 (3)
Range High
Note
. Multi-stream wet-sections only.
Sets the range span in
Chart
and
Bargraph
views.
Range Low
Note
. Single-stream wet-sections.
Sets the range zero in
Chart
and
Bargraph
views.
Stream 1 (3)
Range Low
Note
. Multi-stream wet-sections only.
Sets the range zero in
Chart
and
Bargraph
views.
Filter Type
Selects the filter type:
Off / Min / Max / Average / Rolling Point
.
Off
Filter Time
Sets the filter (input) time (5 to 100 seconds).
5 seconds
Wetting Routine
Enables / disables the wetting routine. The wetting routine is initiated
when a
No Samples
situation is detected. It energizes the high
calibration solution valve for 1 minute every 24 hours.
The wetting routine is terminated if sample returns.
Continued on next page…