Navigator 500
Sodium analyzer
6 Operation overview – transmitter
OI/ASO550–EN Rev. G
29
6 Operation overview – transmitter
6.1 Front panel keys
The transmitter is operated using the keys on the front panel. These enable local navigation and selection of software options on all
displays, acknowledgement and data logging and monitoring. Prompts associated with active keys are displayed on each screen.
Diagnostic messages are detailed in Appendix A.1, page 72, display icon descriptions are detailed in Section 6.10, page 40.
Key functions are described in the following table:
Fig. 6.1 Front panel keys
Key
Function
Description
A
Navigation key – left
and
Operator Level
access key
At menu level, selects the highlighted menu item, edit a selection or return to the previous
menu level.
When
Operator
page is displayed, opens or closes the
Operator
menu.
B
View key
Toggles the view between
Operator
pages,
Diagnostic View
and
Calibration Log
screens.
Note
. Not enabled in
Configuration
mode.
C
Up key
Used to navigate up menu lists, highlight menu items and increase displayed values.
D
Down key
Used to navigate down menu lists, highlight menu items and decrease displayed values.
E
Group key
Toggles between:
Operator
pages (1 to 5) when an
Operator
page is selected at the
Group
key.
View
screens (
Alarms, Outputs, Signals, Chart
and
Diagnostic
) when the
Diagnostic
View
screen is selected at the
Group
key.
Log
screens (
Alarm, Audit, Diagnostic
and
Calibration
) when the
Calibration Log
screen is selected at the
Group
key.
Note
. Not enabled in
Configuration
mode.
F
Navigation key – right
and
Cal
shortcut key
At menu level, selects the highlighted menu item, operation button or edits a selection.
At
Operator
page level, used as a shortcut key to access the
Calibrate
level.