Navigator 500
Sodium analyzer
5 Calibration
OI/ASO550–EN Rev. G
27
14. Press the
key (below the
Continue
prompt).
The next
Process Calibration
page displayed shows an
average of the reading obtained:
A new value can be added at this page.
15. Use the
/
keys to move the cursor to the numeral
to be edited in the
New 000.00
reading on the screen.
To enter the value, press the
key (below the
Next
prompt).
Once the new value has been entered, press the
key
(below the
Continue
prompt).
The process calibration offset is calculated and a page
with a moving time bard is displayed for the duration of the
calculation:
When the calculation is finished, a new page showing the
new reading and process calibration offset values is
displayed
The
Process Calibration (PC) Offset
is displayed in the
transmitter’s
Signals View – see
Section 6.6, page 33.
It is toggled with the
Concentration Value
. The Process
Calibration (PC)
Offset
is also displayed in the transmitter’s
Calibration Log
.
16. Press the
key (below the
Exit
prompt) to return to the
Sensor Calibration
page
17. Press the
key (below the
Back
prompt).
The
Calibrate
page is displayed:
18. Press the
key (below the
Exit
prompt) to exit the
configuration level menu.
Note.
If the new concentration value is not entered
into the transmitter within 5 hours of the sample
collection the process calibration procedure is
aborted automatically.
New 000.00
AVG 2.93 ppb
Process Calibration
Next
Continue
3
New 00 .00
AVG 2.93 ppb
Process Calibration
Next
Continue
AVG 2.93 ppb
Process Calibration
Abort
Note.
The PC Offset can be removed from the
concentration calculation by carrying out a single- or
two-point calibration.
Exit
New
3.00 ppb
PC Offset 0.007
Process Calibration
Sensor Calibration
Process Calibration
Back
Select
Menu
Calibrate
Exit
Select