Navigator 500
Sodium analyzer
5 Calibration
26
OI/ASO550–EN Rev. G
7. Press the
key (below the
Select
prompt).
The
Sensor 1 (4)
/
Sensor Three Stream
page (dependent
on single-stream / multi-stream configuration) is displayed
with the
Sensor Calibration
menu selected:
8. Press the
key (below the
Select
prompt on the
Sensor
1 (4)
/
Sensor Three Stream
page).
The
Sensor Calibration
page is displayed:
Press the
key to highlight the
Process Calibration
menu.
9. Press the
key (below the
Select
prompt).
The
Process Calibration
page is displayed:
At the wet-section:
10. Place the drain tube from the constant-head unit into the
sample vessel (during the sampling period an average
sodium concentration value is determined).
At the transmitter:
11. Press the
key (below the
Yes
prompt).
The
Process Calibration
/
Sample Collection
page is
displayed:
12. Press the
key (below the
Start
prompt).
The next
Process Calibration
/
Sample Collection
page is
displayed with a progress bar:
When sufficient sample has been collected in the sample
vessel, the collected sample must be analyzed using a
laboratory method or another analyzer to determine the
concentration of sodium in the sample.
13. Press the
key (below the
Stop
prompt) to stop the
procedure manually.
The next
Process Calibration
/
Sample Complete
page is
displayed with the progress bar populated:
Sensor 1
Sensor Calibration
Scheduled Calibration
Calibration Setup
Restore Defaults
Back
Select
Sensor Calibration
Low Calibration
High Calibration
Two-Point Cal
Regeneration Cycle
Process Calibration
Grab Sample
Back
Select
Process Calibration
Start Procedure
No
Yes
Note.
To abort the process calibration procedure
during this collection press the
key (below the
Abort
prompt).
Note.
If the procedure is not stopped manually, it
stops automatically after 30 minutes.
Process Calibration
PV 2.93
ppb
Sample Collection
Abort
Start
AVG 2.92 ppb
Sample Collection
Process Calibration
Abort
Stop
AVG 2.93 ppb
Sample Complete
Process Calibration
Abort
Continue