![ABB Navigator 540 Instruction Manual Download Page 33](http://html.mh-extra.com/html/abb/navigator-540/navigator-540_instruction-manual_4465033.webp)
Navigator 540 | Transmitter |
COM/AWT540/PB–EN Rev. C
33
Description
Slot
Index
Object
type
Data type
Bytes
Store
Access
Default
Valid
range
Notes
Transmitter –
Temperature units
120
45
Simple
Unsigned 8
1
D
R/W
Deg.C
13 to 14
Multi-stream sodium only
Concentration units 120
46
Simple
Unsigned 8
1
D
R/W
ppb
7 to 9
Filter type
120
47
Simple
Unsigned 8
1
D
R/W
Off
0 to 4
Filter units
120
48
Simple
Unsigned 8
1
D
R/W
5
5 to 100
Wetting routine
procedure
120
49
Simple
Unsigned 8
1
D
R/W
Disabled
0 to 1
Stream 1 – Flow
measurement
120
50
Simple
Unsigned 8
1
D
R/W
Disabled
0 to 1
Stream 2 – Flow
measurement
120
51
Simple
Unsigned 8
1
D
R/W
Disabled
0 to 1
Stream 3 – Flow
measurement
120
52
Simple
Unsigned 8
1
D
R/W
Disabled
0 to 1
Out of sample
detection
120
53
Simple
Unsigned 8
1
D
R/W
Disabled
0 to 1
Stream sequence –
Position 1
120
54
Simple
Unsigned 8
1
D
R/W
Daily
0 to 6
Stream sequence –
Position 2
120
55
Simple
Unsigned 8
1
D
R/W
Weekly
0 to 7
Stream sequence –
Position 3
120
56
Simple
Unsigned 8
1
D
R/W
Monthly
0 to 5
Stream sequence –
Position 4
120
57
Simple
Unsigned 8
1
D
R/W
1
0 to 3
0 – End Of
Sequence
1 – Stream 1
2 – Stream 2
3 – Stream 3
Stream sequence –
Position 5
120
58
Simple
Unsigned 8
1
D
R/W
2
Stream sequence –
Position 6
120
59
Simple
Unsigned 8
1
D
R/W
1
Stream sequence –
Position 7
120
60
Simple
Unsigned 8
1
D
R/W
2
Stream sequence –
Position 8
120
61
Simple
Unsigned 8
1
D
R/W
1
Stream 1 – Enable
120
62
Simple
Unsigned 8
1
D
R/W
2
Stream 2 – Enable
120
63
Simple
Unsigned 8
1
D
R/W
1
Stream 3 – Enable
120
64
Simple
Unsigned 8
1
D
R/W
2
Scheduled
calibration type
120
65
Simple
Unsigned 8
1
D
R/W
Scheduled
calibration
disabled
0 to 5
Scheduled
calibration
frequency units
120
66
Simple
Unsigned 8
1
D
R/W
Frequency
off
0 to 3
Scheduled
calibration
frequency daily
120
67
Simple
Unsigned 8
1
D
R/W
Daily
0 to 6
Scheduled
calibration
frequency weekly
120
68
Simple
Unsigned 8
1
D
R/W
Weekly
0 to 7
Scheduled
calibration
frequency monthly
120
69
Simple
Unsigned 8
1
D
R/W
Monthly
0 to 5
Table A.26: Multi-stream sensors – units, filter and scheduled calibration configuration