Navigator 540 | Transmitter |
COM/AWT540/PB–EN Rev. C
11
The ABB GSD file (Ident No. 3402) is divided into 2 sections:
—
General specifications
Identification of the device, together with hardware and
software versions, baud rates supported and the possible
time intervals for monitoring times.
—
DP Slave-related specifications
Information about the user parameter block for
device-specific configuration and modules containing
details of the input and output data that can be exchanged
cyclically with a PROFIBUS master.
The Navigator AWT540 GSD file (ABB_3402.gsd) is available
for download from the ABB website at: www.abb.com/fieldbus
(follow the link for PROFIBUS DP field devices).
4.6.3 Configuration from the Navigator AWT540 transmitter
PROFIBUS-related parameters are configured from the
Navigator AWT540’s
Communication / PROFIBUS
page – see
Section 3, page 7 for access details.
PROFIBUS
parameters are readable at all access levels. Write
access to secure (writable) parameters is possible only at
Advanced
level. Refer to associated Operating instructions
(OI/ASO550-EN [sodium], OI/ADS550-EN [dissolved oxygen,
OI/AHM550-EN [hydrazine]) for details of
Access Level
page
parameters.
PROFIBUS
parameters options are described in Table 4.6.
4.7 Navigator AWT540-DP PROFIBUS datasheet
Parameter Description
Station Address
Sets the PROFIBUS
Station Address
value
–
see Section 4.6.1, page 10
.
Baud Rate
Indicates the baud rate (in kbps) the
transmitter has locked on to (read-only
parameter).
Table 4.6: PROFIBUS parameter descriptions
Item
Detail
PROFIBUS device
name
NAVIGATOR 500
(product: Navigator AWT540)
Applicable
standards
IEC61158 (Type 3)
IEC61784 (CPF3/1)
Protocols
supported
PROFIBUS-DP (DPV0)
PROFIBUS-DP extensions
(DPV1 Class 1 supported)
Note
. DPV1 Class 2 is not supported
Profile ident
support
Standard: 9700, 9740, 9741
Device specific: 3431 (4xAI, 2xTOT, 1xAO)
PROFIBUS unit
type
PROFIBUS DPV1 Slave
PROFIBUS media
type
RS485 (EIA-485), galvanically isolated
PROFIBUS bus
connection
Wiring terminals A1/B1 (in) and A2/B2
(out)
PROFIBUS slot /
index
PROFIBUS commands must include
length (in bytes) as well as slot / index.
This is so that multiple values from the
same slot (except strings) can be read /
written in a single transaction (up to the
maximum of 240 bytes). The length must
be a multiple of the 'Bytes value for the
required slot.
The index specifies the starting index for
the values returned. The starting index
plus the number of bytes must not go
beyond the end of the table. Strings must
also specify the correct length in bytes but
only single strings can be accessed in one
transaction.
Bus address
range
1 to 126 via local display interface
1 to 125 via Set_Slave_Addr service
Baud rates
supported
9.6 kbps
19.2 kbps
45.45 kbps
93.75 kbps
187.5 kbps
500 kbps
1.5 Mbps
3 Mbps
6 Mbps
12 Mbps
Master Class 1 –
Slave cyclic
services (MS0)
Set_Prm
Chk_Cfg
Set_Slave_Addr
Data_Exchange
Get_Diag
Get_Cfg
Rd_Inp
Rd_Outp
Master Class 1 –
Slave acyclic
services (MS1)
MS1_Read
MS1_Write
Device-specific
GSD file
ABB_3402.gsd
Configuration
support
Local display interface
Table 4.7: PROFIBUS datasheet