30
COM/AWT540/PB–EN Rev. C
| Navigator 540 | Transmitter
Sensor 2 –
Concentration units
120
12
Simple
Unsigned 8
1
D
R/W
ppb
7 to 9
7 – ppb
8 – ug/l
9 – ug/Kg
Sensor 2 –
Filter type
120
13
Simple
Unsigned 8
1
D
R/W
Off
0 to 4
0 – Off
1 – Minimum value
2 – Maximum value
3 – Average value
4 – Sliding average
Sensor 2 –
Filter units
120
14
Simple
Unsigned 8
1
D
R/W
5
5 to 100
Samples / seconds
Sensor 2 – Wetting
routine procedure
120
15
Simple
Unsigned 8
1
D
R/W
Disabled
0 to 1
0 – Disabled
1 – Enabled
Sensor 2 – Flow
measurement
120
16
Simple
Unsigned 8
1
D
R/W
Disabled
0 to 1
0 – Disabled
1 – Enabled
Sensor 2 – Out of
sample detection
120
17
Simple
Unsigned 8
1
D
R/W
Disabled
0 to 1
0 – Disabled
1 – Enabled
Sensor 2 –
Scheduled
Calibration
Type
120
18
Simple
Unsigned 8
1
D
R/W
Scheduled
calibration
disabled
0 to 5
0 – Scheduled calibration disabled
1 – One point calibration
(low standard)
2 – One point calibration
(high standard)
3 – Two point calibration
4 – Regene two point
calibration
5 – One point air calibration
Sensor 2 –
Scheduled
calibration
frequency units
120
19
Simple
Unsigned 8
1
D
R/W
Frequency
off
0 to 3
0 – Frequency off
1 – Daily
2 – Weekly
3 – Monthly
Sensor 2 –
Scheduled
calibration
frequency daily
120
20
Simple
Unsigned 8
1
D
R/W
Daily
0 to 6
0 – Daily
1 – Two days
2 – Three days
3 – Four days
4 – Five days
5 – Six days
6 – Seven days
Sensor 2 –
Scheduled
calibration
frequency weekly
120
21
Simple
Unsigned 8
1
D
R/W
Weekly
0 to 7
0 – Weekly
1 – Two weeks
2 – Three weeks
3 – Four weeks
4 – Five weeks
5 – Six weeks
6 – Seven weeks
7 – Eight weeks
Sensor 2 –
Scheduled
calibration
frequency monthly
120
22
Simple
Unsigned 8
1
D
R/W
Monthly
0 to 5
0 – Monthly
1 – Two months
2 – Three months
3 – Four months
4 – Six months
5 – Twelve months
Description
Slot
Index
Object
type
Data type
Bytes
Store
Access
Default
Valid
range
Notes
Table A.25: Single-stream sensors – units, filter and scheduled calibration configuration (continued)