![ABB LWT300 series User Manual Download Page 61](http://html.mh-extra.com/html/abb/lwt300-series/lwt300-series_user-manual_4452061.webp)
Configuring the instrument
57
Setting the vessel diameter
To set the vessel diameter:
1
Select
Device Setup
>
Linearization
>
Volume Calculation
>
Cylinder Diameter
.
2
Press
Edit
.
3
Set the vessel diameter and press
OK
.
Setting the vessel length
To set the vessel length:
1
Select
Device Setup
>
Linearization
>
Volume Calculation
>
Cylinder Length
.
2
Press
Edit
.
3
Set the vessel length and press
OK
.
Generating a linearization table automatically
Once you have set your vessel parameters, you can generate a linearization table automatically by
selectring
Device Setup
>
Linearization
>
Volume Calculation
>
Gen Linearization
and clicking
OK
.
This automatically create a linearization table based on your vessel parameters.
Setting up linearization points
The first point is always the point with the lowest value and all other points must be set in increasing order.
To properly configure the linearization feature:
1
Select
Device Setup
>
Linearization
>
Set LIN. Points
.
The
00
linearization point is highlighted (see below).
2
Press the Left action key to select the measured level (
In
).
3
Press the Right action key to edit the
In
value, and press
OK
when you are done.
4
Press the Left action key to select the output associated with the measured level (
Out
).
5
Press the Right action key to edit the
Out
value, and press
OK
when you are done.
6
Press the Left action key to highlight the linearization point again.
7
With the Up and Down arrow keys, change the linearization point.
8
Repeat steps 2 to 7 for all the required linearization points.
9
All your linearization points that you just created compose your linearization table.
NOTICE
If you need to modify any linearization point or the maximum value in the future, you
MUST
first disable the linearization function (see “Enabling the linearization function” on page 58)
or use the automatically-generated table (see “Generating a linearization table automatically”
on page 57).
Summary of Contents for LWT300 series
Page 1: ... USER GUIDE LWT300 series Guided wave radar level transmitter Modbus ...
Page 36: ...Page intentionally left blank ...
Page 40: ...Page intentionally left blank ...
Page 66: ...Page intentionally left blank ...
Page 70: ...Page intentionally left blank ...
Page 76: ...Page intentionally left blank ...
Page 96: ...Page intentionally left blank ...
Page 98: ...Page intentionally left blank ...
Page 104: ...Page intentionally left blank ...
Page 105: ...Page intentionally left blank ...
Page 106: ...Page intentionally left blank ...
Page 107: ......