D184B122U02
S4
25
Submenu/
Parameter
Submenu/
Parameter
Submenu/
Parameter
Selections
Value Range
Comments
Prog Level
PP-Code
Old PP-Code:
1–9999
Normally the Prog. Protection Code is equal
to "0". If the Prog. Prot. is selected, the
question "New PP Code?" is displayed. If a
code is entered between (1 and 9999) then
an attempt to access the PP Code protect-
ed menus by an operator will result in the
display of the message
** PP Code ? ** indicating that the operator
must enter the correct code before access
to the menus is granted. To delete the
PP Code it is necessary to access the
submenu "PP Code". After pressing
ENTER the "Old PP-Code?" is requested.
After entering the correct value it is possible
in the next submenu "New PP-Code?" to
enter a new PP-Code value or by entering
"0" to turn off the program protection.
ENTER
0
New PP-Code
1–9999
0
Submenu
Language
Unit Qv
Unit Tv
Qmax
Damping 1Tau
Display mode
Display 1st line
Display 2nd line
Display 3rd line
Display 4th line
Iout type
Iout Low Alarm
Iout High Alarm
Pulse factor
Pulse width
With EASY -SETUP it is possible to config-
ure the metering system for operation using
the fewest possible steps. The menus listed
to the right are displayed for entry one after
the other in "EASY-SETUP". If the selections
1 big, 1 small are selected then the selection
menus Display 3rd line, Display 4th line are
not displayed. For HART instruments the
selection “Iout type” is not displayed. Using
ENTER the parameters can be changed one
after the other, or if they are not to be
changed, the C/CE key can be used to pro-
ceed to the next
parameter. After making an entry with
ENTER the entry can still be changed if
necessary by using the ENTER Function
and then using the STEP-key to proceed to
the next parameter. For numeric entries, the
minimum and maximum entry values are
also displayed.
EASY SETUP
EASY SETUP
ENTER
Language
German
English
French
Finish
Spanish
Italian
Dutch
Danish
Swedish
The languages available for the menu
display are shown in the “Language”
submenu.
Menus in the Service Code section use the
English language.
English
Language
ENTER
English
Submenu
Operating mode
Operating mode
Standard
Fast
Piston Pump
"Standard": selection used for continuous
flow metering.
"Fast": accelerated measured value pro-
cessing (e.g. short batch operation"
“Piston Pump”: for pulsating flows. In this
operating mode improved reproducibility of
the measurements is achieved for pulsating
flow in piston pump operation. The
totalizer is automatically set for the operat-
ing mode Difference Totalizer. Any reverse
flow which may occur is automatically de-
ducted from the totalizer value.
ENTER
Standard