Parameterization
82 FSM4000
D184B140U02
Submenu/parameter
Value range
Comment
Submenu
Alert Elec Bal
Set the thresholds and alarm behavior for the electrode
balance.
The electrode balance is calculated as the ratio between
the electrode AC voltages E1/E2. This value is displayed
in [%]. If the electrode voltage E2 is larger than E1, the
value is assigned a minus sign and the ratio E2/E1 is
displayed.
Min. Val
100 ... 300
Lower range limit for alarm.
0.0 %
Max. Val
100 ... 300
Upper range limit for alarm.
300.0 %
Status signal
Defines the alarm behavior.
Off
Off, warning, error
Note:
To properly display the electrode voltage, the
electrode zero must be adjusted. For additional
information, see "Submenu Adjustment" on page 85.
Submenu
Alert Mag Lin
Set the thresholds and alarm behavior for the magnetic
linearity.
Min. Val
0 ... 100
Lower range limit for alarm.
0.0 %
Max. Val
0 ... 100
Upper range limit for alarm.
100.0 %
Status signal
Defines the alarm behavior.
Off
Off, warning, error
Note:
If magnetic linearity is being measured (at least
every 5 minutes), the last valid flowrate value during the
measurement is retained (approx. 2 s).
Submenu
Alert SNR
Set the thresholds and alarm behavior for the signal-to-
noise ratio.
Min. Val
0 ... 100
Lower range limit for alarm.
0.0 %
Max. Val
0 ... 100
Upper range limit for alarm.
100.0 %
Status signal
Defines the alarm behavior.
Off
Off, warning, error