Parameterization
D184B140U02 FSM4000
87
Submenu/parameter
Value range
Comment
Submenu
Syst. Adjust
Start
System
Zero
0.0
%
Manual
Autom. Calibration
Check zero, if necessary (when adapting an older
flowmeter sensor). Manual entry, e.g., when replacing the
transmitter.
System
zero
point
0.0
%
Numeric entry
Automatic adjustment: Valve must be closed, fluid in
flowmeter sensor must be at absolute standstill. Press
ENTER to perform the automatic adjustment. The limit for
the zero point is ± 10%. If the value is outside this limit, no
calibration is performed.
FSM4000
01/2008 Display
only
D699G004U01
C.10
Display software version. 01.2008 specifies the date of
the release. C.10 = Version level, for standard and HART
protocol models.
FSM4000
PA3.0
Display
only
D699G004U02
C.10
For PROFIBUS PA software.
FSM4000
FF
Display
only
D699G004U03
C.10
For FOUNDATION Fieldbus communication software.