Parameterization
62 FSM4000
D184B140U02
Pos: 33.8 /Konfiguration, Parametrierung/Durchfluss/FSM4000/Parametrierung/Parameterübersicht FSM4000, Teil 1 @ 16\mod_1196333152656_3101.doc @ 143587
Submenu/parameter
Value range
Comment
Prog.
Level
Locked
"Locked": All entries in the standard menus can be read
Prog.
Level
All "Standard" menus are shaded in gray
Locked
Prog.
Level
Standard
Prog.
Level
Specialist
Prog.
Level
Service
"Standard": Contains all menu items needed to operate
the device.
"Specialist": Contains all menu items accessible by the
customer.
"Service": Additional menus can be displayed by entering
the correct service code number (for ABB Service only).
Servicecode?
0
Prog.
Level
Pg.prot.code
Old
Pg.prot.code:
1 ... 9999
0
New
Pg.prot.code
1 ... 9999
0
The Prog. Protection Code is normally "0". If Prog.
Protection is selected, the question "New PP code?" is
displayed. If the user enters an incorrect code between 1
... 9999, a message **Prog.Prot.** appears, indicating
that the menu is locked and requires the correct PP code.
To suspend the PP-Code function, you must access the
"PP-Code" menu. Select ENTER and you will be asked
for the old PP code. You must enter the correct old code
before the message "New PP-Code?" is displayed. Enter
a new code or select "0" to disable the PP-Code function.
Submenu
EASY
SETUP
EASY
SETUP
Language
Unit Qv
Unit Zv
Qmax
Damping 1,000
Display mode
Display line 1
Display line 2
Display line 3
Display line 4
Iout Art
Iout Alarm
Low Alarm
High Alarm
Pulse factor Vol.
Pulse width
The EASY SETUP function allows you to configure the
system in an easy manner for operation. Menus are
displayed in sequential order by EASY SETUP. If the
display mode 1 big, 1 small is selected, the selection
menus for Display 3rd line and Display 4th line are not
displayed. For HART devices, the selection menu "lout
type" is also not displayed. The ENTER function allows
you to change parameters; the C/CE button jumps to the
next entry. Press ENTER again to change parameters
and, if necessary, you can call the next parameter via the
STEP button. For numeric entries, the minimum and
maximum settings are displayed.