![ABB CoriolisMaster FCB400 Manual Download Page 84](http://html1.mh-extra.com/html/abb/coriolismaster-fcb400/coriolismaster-fcb400_manual_3662119084.webp)
84
CoriolisMaster FCB400, FCH400
CORIOLIS MASS FLOWMETER | CI/FCB400/FCH400-EN REV. H
… 8 Operation
… Selecting and changing parameters
Exiting the setup
For some menu items, values must be entered. If you don't want
to change the parameter, you can exit the menu as described
below.
1. Pressing
(Next) repeatedly moves the cursor to the right.
Once the cursor reaches the end position, ‘Cancel’ is
displayed in the lower right of the screen.
2.
terminates editing and exits the menu item. Use
to
return to the start.
Note
The LCD display automatically returns to the process display
three minutes after the last button has been actuated.
Error messages on the LCD display
In the event of an error, a message consisting of a symbol and
text (e.g. Electronics) appears at the bottom of the process
screen.
The text displayed provides information about the area in which
the error has occurred.
Process display
Electronics
The error messages are divided into four groups in accordance
with the NAMUR classification scheme. The group assignment
can only be changed using a DTM or EDD:
Symbol Description
Error / failure
Function check
Outside of the specification
Maintenance required
The error messages are also divided into the following areas:
Range Description
Operation
Error / alarm due to the current operating
conditions.
Sensor
Error / alarm of the flowmeter sensor.
Electronics
Error / alarm of the electronics.
Configuration
Error / alarm due to device configuration.
Note
For a detailed description of errors and information regarding
troubleshooting, refer to the chapter titled "Diagnosis / Error
messages" in the operating instruction.