CoriolisMaster FCB400, FCH400
CORIOLIS MASS FLOWMETER | CI/FCB400/FCH400-EN REV. H
19
Inputs and outputs with optional plug-in cards
Operating values (general)
Type of protection – ‘nA’ / ‘NI’
U
N
I
N
U
N
I
N
Current output V3 / V4, active*
Terminals V3 / V4 and V1 / V2*
30 V
30 mA
30 V
30 mA
Current output V1 / V2, passive**
Current output V3 / V4, passive**
Terminals V1 / V2** or V3 / V4**
30 V
30 mA
30 V
30 mA
Digital output V3 / V4, active*
Terminals V3 / V4 and V1 / V2*
30 V
25 mA
30 V
25 mA
Digital output V1 / V2, passive**
Digital output V3 / V4, passive**
Terminals V1 / V2** or V3 / V4**
30 V
30 mA
30 V
30 mA
Digital input V3 / V4, active*
Terminals V3 / V4 and V1 / V2
30 V
3,45 mA
30 V
3,45 mA
Digital input V1 / V2, passive**
Digital input V3 / V4, passive**
Terminals V1 / V2** or V3 / V4**
30 V
3,45 mA
30 V
3,45 mA
Digital output V1 / V2, active*
Digital output V3 / V4, active*
Terminals V1 / V2** or V3 / V4**
24 V
22,5 mA
30 V
30 mA
Modbus® / PROFIBUS DP®
Terminals V1 / V2
30 V
30 mA
30 V
30 mA
Ethernet card
Ethernet (various protocols)
Port 1 / Port 2
57 V
417 mA
57 V
417 mA
Ethernet card in connection with Power over Ethernet (POE card)
Ethernet (various protocols)
Port 1 / Port 2
57 V
417 mA
57 V
417 mA
* Only in conjunction with additional ‘24 V DC loop power supply (blue)’ plug-in card in slot OC1.
** The terminal assignment depends on the model number or the slot assignments. For connection examples, see
Connection examples
on page 53.