74
AW T4 20 |
U N I V ER S A L 4 - W I R E , D UA L- I N P U T T R A N SM I T T ER | O I/AW T4 2 0 - EN R E V. B
…15 Calibration procedures
…Turbidity TSS calibration
2-Point calibration
1
At the
Calibrate
level, press the key:
The
Calibrate
menu is displayed:
2
Use the
/
keys to select the sensor to be calibrated,
and press the key to confirm selection.
The
Turbidity Cal
menu is displayed:
3
Press the
/
keys to select
Turbidity Cal
and
press the key to confirm selection.
The
Calibration Type
is displayed:
4
Use the key to edit the
Calibration Type
.
Use the
/
keys to select the required
Calibration Type
and press the key to confirm selection.
Press the and key to proceed to the next step.
The
Low Solution Value
is displayed:
5
Use the key to edit the
solution value
.
Use the
/
keys to set the value and press the key
to confirm changes.
Press the and key to proceed to the next step.
The
High Solution Value
is displayed:
6
Use the key to edit the solution value .
Use the
/
keys to set the value and press the key
to confirm changes.
Press the and key to proceed to the next step.
Calibrate
Back
Select
S1 : Turb TSS
S2 : Turb TSS
Hold Outputs
Calibrate
Back
Select
Sensor Verification
Turbidity Cal
TSS Cal
TSS Manual Cal
Sample Collection
Collection Complete
Turbidity Cal
Calibration Type
2-Point
Next
Edit
Turbidity Cal
Low Solution Value
0.000 NTU
Next
Edit
Turbidity Cal
High Solution Value
20000 NTU
Next
Edit
Calibrate
Start Cal ?
Place in Low Sol
0.000 NTU
Abort
Continue