…15 Calibration procedures
2-electrode conductivity, resistivity or concentration calibration
Once the sensor has been installed and has reached the
temperature of the process solution, verify the process variable
value using a grab sample and an external validation device
having the same type of temperature compensation.
1
At the
Calibrate
level, press the key:
The
Calibrate
menu is displayed:
2
Use the
/
keys to select
S1 : TC
and
press the key.
The
S1 : TC
menu is displayed:
3
Use the
/
keys to select
Conductivity Cal
and press the .
The
Conductivity Cal
menu is displayed:
4
Confirm the displayed reading is stable and the key.
5
Press the key to enter a new value (the transmitter
takes several seconds to validate the calibration):
Invalid new calibration values generate an error
message and the calibration value is not accepted.
If the new value is valid,
Slope
and
Offset
values
are displayed.
Calibrate
Exit
Select
S1 : TC
Hold Outputs
S1 : TC
Exit
Select
Conductivity Cal
Temperature Cal
Edit Cal
Restore Cal Defaults
Conductivity Cal
PV 24929
µ
S/cm
Continue When Stable
Abort
Continue
Conductivity Cal
PV 24929
µ
S/cm
New 02495
Exit
Continue
Conductivity Cal
PV 24923
µ
S/cm
Abort
Conductivity Cal
PV 24935
µ
S/cm
Calibration Failed
Offset Too High
Exit