52
AO2000-LS25
LASER ANALYZERS | OI/AO2000-LS25-EN REV. D
… 10 Configuration, parameterization
Software start-up
After starting the program, a dialog for selecting the
communication link is shown:
Figure 35: Interface selection
The version of the service program is printed in the right bottom
corner (in the example above V.1.2.5.1).
Serial connection
For direct serial or modem connection, select the PC serial port
from the dropdown list of available ports and press the
corresponding button.
Modem connection
For a modem connection, the phone number to be dialed must
be entered.
If the first digit in this phone number is a “0”, the software will
pause after dialing the zero before dialing the rest of the digits.
If additional pausing is needed, apply "," between the digits.
LAN connection
If the communication is used via LAN, the IP address and port
number must be entered.
• The IP address and port number are preset to
192.168.1.237 and 5001 at the factory.
If the IP address and port number are to be changed, the gas
analyzer must be connected via the serial connection and the
TCP/IP parameters must be configured according to
modem configuration <TCP/IP & modem configuration>
page 64.
Selecting configuration mode
After connecting to the device, the configuration mode must be
selected:
• <User mode> (User mode)
• <Advanced mode> (Advanced mode, unrestricted access
to all device parameters)
• <Demo mode> (No connection to device)
Note
• If another LS25 gas analyzer is to be configured, the service
program must be closed and opened again.
• It is possible to open several program windows
simultaneously to communicate with different devices with
one PC.
<User mode> (User mode)
This User mode provides a simplified interface with limited
access to the device setting parameters and should be used
during installation and normal operation of the device.
<Advanced mode> (Advanced mode)
A password is only required for Advanced mode that enables full
access to all the instrument setting parameters.
Extended mode should be used during service adjustment of the
device. For the necessary password, please contact ABB Service.
<Demo mode> (Demo mode)
The service program can also be started without a connection to
a device by pressing <Demo mode>. This is for demonstration
purposes of the service program.
Two special files are required for this purpose:
*.dmp, or *.rea and *.set, or *.stt
These files can be created using the service program when the
PC is connected to a device.
The files are created by downloading the device readings to *.rea
and settings to *.set (see
Alternatively, the files supplied on the data storage medium can
be used. These files were downloaded from the instrument at
the factory prior to shipment.