![ABB AO2000-LS25 Manual Download Page 39](http://html.mh-extra.com/html/abb/ao2000-ls25/ao2000-ls25_manual_4322039.webp)
AO2000-LS25
LASER ANALYZERS | OI/AO2000-LS25-EN REV. D
39
8
Electrical connections
Safety instructions
WARNING
Risk of injury due to live parts.
Improper work on the electrical connections can result in
electric shock.
• Connect the device only with the power supply switched
off.
• Observe the applicable standards and regulations for the
electrical connection.
Note
When using the device in potentially explosive atmospheres,
note the additional data in
atmospheres in accordance with ATEX and IECEx
Use in hazardous areas in accordance with CSA
Note
• The polarities listed in the column “Description” are used for
naming only and do not necessarily correspond to the
polarities of the applied voltages.
• All potentials are floating and none of them should be
grounded to box. This applies to all connection tables.
Transmitter unit interface
Figure 23: Electrical connections on the transmitter unit (viewed from below)
Pos.
Description
1
Connector for receiver unit (signal cable)
Power supply and signals from and to the receiver unit.
Connection assignment, see
2
Network connection
10 or 10/100 Base-T Ethernet (RJ45).
Connection assignment, see
3
Connection plug for power supply and analog inputs
Power supply and signals from the external sensors.
Connection assignment, see
on
page 41.
4
Service connection
RS232 interface (galvanically isolated).
Connection assignment, see