AO2000-LS25
LASER ANALYZERS | OI/AO2000-LS25-EN REV. D
51
4. Enter the IP address of the laser analyzer and press <ENTER>
to confirm. The factory-set IP address is 192.168.1.237.
• In the system modules list the laser analyzer will be
shown with a “New” status.
10
Configuration, parameterization
Parameterization of the device
The instrument calculates the gas concentration from the
measured signal, which depends on several process parameters.
The instrument must therefore be configured for the specific
installation.
With the service program the user communicates with the
instrument and changes installation or process relevant
settings.
With a PC connected to the instrument via a standard serial
communication port the operator can perform the following
major tasks:
• Monitor the measured concentration and the laser beam
transmission;
• Plot the measured concentration as a function of time;
• Log the concentration and other control parameters to a file;
• Display the measured signal;
• Save the measured signal and other control parameters to a
file;
• Observe and erase the instrument error logging;
• Configure the 4 to 20 mA process temperature and pressure
inputs;
• Change the measurement averaging time;
• Set the optical path parameters;
• Change the concentration unit;
• Change the format of the displayed concentration;
• Change the instrument clock time;
• Calibrate the instrument;
• Save all instrument parameters to a file;
• Restore the instrument parameters from a file.
• Configure the Ethernet connection
Through different menus the necessary installation parameters
may be set.
After setting the necessary parameters the PC is no longer
needed. The analyzer has all the parameters stored in the
internal memory.
The PC can therefore be disconnected and the analyzer can be
turned off and on without resetting the parameters.