2600T Series Pressure transmitters |
OI/266/HART-EN Rev. D 27
Pressure or absolute pressure measurement of a liquid in a pipe
– Place the tap at the side of the line.
– Mount the transmitter beside or below the tap for clean fluids, above the tap for dirty fluids.
– Connect the + (H) side of the transmitter to the pipe.
Pressure or absolute pressure measurement of a condensable vapor in a pipe
– Place the tap at the side of the line.
– Mount the transmitter below the tap.
– Connect the + (H) side of the transmitter to the pipe.
– Fill the vertical section of the connecting line to the tap with a compatible liquid through the dedicated filling tee.
Figure 32: Gauge or absolute pressure measurement of a liquid in a pipe with DP and P style transmitter
Figure 33: Gauge or absolute pressure measurement of condensable vapor in a pipe with DP and P style transmitter (filling tee installed)
Dirty fluids
Dirty fluids
Clean fluids
Clean fluids
Summary of Contents for 266 HART Series
Page 40: ...40 OI 266 HART EN Rev D 2600T Series Pressure transmitters 266 Models HART Device Set up ...
Page 41: ...2600T Series Pressure transmitters OI 266 HART EN Rev D 41 ...
Page 42: ...42 OI 266 HART EN Rev D 2600T Series Pressure transmitters 266 Models HART ...
Page 43: ...2600T Series Pressure transmitters OI 266 HART EN Rev D 43 ...
Page 44: ...44 OI 266 HART EN Rev D 2600T Series Pressure transmitters 266 Models HART Display ...
Page 45: ...2600T Series Pressure transmitters OI 266 HART EN Rev D 45 ...
Page 46: ...46 OI 266 HART EN Rev D 2600T Series Pressure transmitters 266 Models HART ...
Page 48: ...48 OI 266 HART EN Rev D 2600T Series Pressure transmitters 266 Models HART Calibrate ...
Page 49: ...2600T Series Pressure transmitters OI 266 HART EN Rev D 49 Totalizer ...
Page 50: ...50 OI 266 HART EN Rev D 2600T Series Pressure transmitters 266 Models HART ...
Page 51: ...2600T Series Pressure transmitters OI 266 HART EN Rev D 51 ...
Page 52: ...52 OI 266 HART EN Rev D 2600T Series Pressure transmitters 266 Models HART Diagnostics ...
Page 53: ...2600T Series Pressure transmitters OI 266 HART EN Rev D 53 Device Info ...
Page 54: ...54 OI 266 HART EN Rev D 2600T Series Pressure transmitters 266 Models HART Communication ...
Page 77: ...2600T Series Pressure transmitters OI 266 HART EN Rev D 77 ...
Page 78: ...78 OI 266 HART EN Rev D 2600T Series Pressure transmitters 266 Models HART ...
Page 79: ...2600T Series Pressure transmitters OI 266 HART EN Rev D 79 ...