6.1.5 Control light to indicate malfunction
Af ter tur ning on the s-max sel la the green con trol light
lights up (see chap ter
5.5) and re mains lit, which in di ca tes that the stair clim bing sys tem is in or -
der and rea dy to use.
Po ten ti al mal functi on is in di ca ted by a light di ode. Af ter a pe ri od of 3 se -
conds each time a blin king sig nal is re pea ted.
In case of mal functi on, turn the de vi ce first off and then on again. If the re is
still an in di ca ti on of mal functi on, char ge the bat te ries in case of low bat te ry
char ge. For all ot her ca ses of mal functi on, plea se call your aut ho ri zed dea -
ler.
6.2
Single step mode
The sing le step switch
is lo ca ted on the clim bing unit.
Plea se ac ti va te the sing le step mode to prac ti xe and in par ti cu lar to climb
win ding stairs.
If you ac ti va te the UP/DOWN switch in that mode, the de vi ce will climb only
one step at a time. The clim bing sys tem stops in the cor rect po si ti on, even if
you press the UP/DOWN switch con ti nu ous ly.
The clim bing me cha nism can not un in ten tio nal ly be ac ti vat ed. You can take
you time to move the main wheels to the next step's edge.
To con ti nue clim bing you sim ply re lea se the UP/DOWN switch and then
press it again.
6.3
Climber
The clim ber
(see pic tu re) must be pul led in to top most po si ti on by me ans
of the UP/DOWN switch if you want to:
•
dri ve on le vel ground and on slo pes
•
check the sa fe ty bra kes
•
be gin clim bing up stairs or downs tairs
Ho we ver, you need to fold out the clim ber by ac ti va ting the UP/DOWN
switch, di rec ti on up, whe ne ver a per son ta kes a seat in the s-max sella, so
the de vi ce is sup por ted by the clim ber (plea se also see chap ter 6.5).
21
s
-max
sel la
· 160 kg
LED is blinking
Identified malfunction
1 x
overload of the device
2 x
battery pack, undervoltage
3 x
overvoltage
4 x
UP/DOWN switch or its supply line is damaged
5 x
defective electromagnetic brake
6 x
malfunction of electronics