TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
page 39 of 93
6.1.2 Procedure for Oxygen Optode 4835
Procedure for changing the sensor foil using kit 5551:
•
The sensor foil is changed by unscrewing the 2 torx screws in the securing plate, refer Figure
6-4Remove the securing plate and the old foil.
o
Rinse clean the sapphire window with DI water
o
If the removed foil will be used in the future it should be pre-cleaned and rinsed then packed in a
light blocking package marked with the foil type and batch number.
•
Center the new foil to fit the optical window. It is important that the standard foil is mounted with the
black side out.
•
Remount the securing plate.
•
Control and if necessary update the sensing foil coefficients according to the foil certificate, refer
chapters:3.2;4.1;5.1
•
Recalibrate the sensor, refer chapter 6.2.2.
Figure 6-5 Illustration of parts, Optode 4835
Figure 6-4 Removing the securing plate and change the sensing foil.
Securing Plate
Screws
Sensing Foil