TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
page 14 of 93
PhaseCoef
Float
4
Linearization coefficients for calculating
compensated phase
FoilID
String
9
Sensing Foil Identifier
FoilCoefA
Float
14
Foil coefficients, general curve fit function, set A
FoilCoefB
Float
14
Foil coefficients, general curve fit function, set B
FoilPolyDegT
INT
28
Exponents for temperature, general curve fit
function
FoilPolyDegO
INT
28
Exponents for oxygen, general curve fit function
SVUFoilCoef
Float
7
Foil coefficients for the ‘Stern Volmer Uchida’
formula
ConcCoef
Float
2
Linear adjustments coefficients for final O
2
concentration calculation, nominal values 0 (offset)
and 1 (slope).
NomAirPress
Float
1
Nominal air pressure for use in O
2
concentration
calculations
NomAirMix
Float
1
Nominal O
2
percentage in air for use in O
2
concentration calculations
CalDataSat
Float
2
Two point calibration data, raw phase and
temperature @ 100% air saturation
CalDataAPress
Float
1
Two point calibration data, air pressure (hPa)
CalDataZero
Float
2
Two point calibration data, raw phase and
temperature @ 0% air saturation
Enable
RedReference
BOOL
1
Controls the use of the red reference LED
RedReference
Interval
INT
1
Sample interval divisor for use of red reference.
Examples: Value 1 for using red reference for each
sample. Value 10 for using red reference for each
10
th
sample.
Mode
ENUM
1
Operation Mode: ‘AiCaP’, ‘Smart Sensor
Terminal’, ‘AADI Real-Time’, ‘Analog Output’
see chapter 5.7
SC
Low
Enable Sleep
BOOL
1
Enable sleep mode
SC
Low
Enable Polled Mode
BOOL
1
Enable Polled Mode (for RS232), when set to ‘no’
the sensor will sample at the interval given by the
Interval
property, when set to ‘yes’ the sensor will
wait for the ‘Do Sample’ command.
Enable Text
BOOL
1
Enable text, when set to ‘no’ the start up info and
the parameter text is removed
Enable
Decimalformat
BOOL
1
Controls the use of decimal format in the
output string
Analog TempLimit
Float
2
Lower and upper ranger limits for analog
temperature output (Output 2), default -5 to 35°C
*2)
Analog Output1
ENUM
Controls which parameter is presented at analog
Output 1; O2Concentration, AirSaturation,
CalPhase, Fixed1, Fixed2
2)