TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
page 37 of 93
6.1 Changing the Sensor Foil
If the sensor foil gets damaged it can easily be changed. For the 4330 and the 4831 models it is possible
to change between the standard foil and fast response foil. The
Sensor Foil Kit 4733
for Optode 4330,
4831 and 4835 and Semnsor Foil Kit 5551 for Optode 5551 contains 1 standard foils while
Sensor Foil Kit
4794
for Optode 4330 and 4831
contains 1 fast response foils. The content of Kit 4733/4794/5551 is
given in Table 6-1, and a procedure for changing the foil is given below the table.
NOTE! If you use a foil from a different batch, new foil coefficients must be entered.
NOTE! After changing the foil and entering new foil coefficients a new two or multipoint calibration should
be performed (see below).
Table 6-1 Contents of Sensor Foil Kit 4733/4794/5551
Part no.
Pieces
Description
Foil Kit
4733
Foil Kit
4794
Foil Kit
5551
1206005A
1
Standard Sensing Foil packed in
aluminium foil
X
1206020
1
Fast Response Sensing Foil packed
in aluminium foil
X
1206019
1
Standard Sensing Foil packed in
aluminium foil
X
1913042
1
Torx key no. T9
X
X
X
1642223
4
M3 x 6mm screw torx a4 Din 965a
(4330/4831)
X
X
X
1642222
2
M2.5 x 6mm screw torx a4 Din 965a
(4835)
X
X
Form No. 721
Calibration Sheet for Sensing Foil (each batch of
foils is calibrated)
X
X