TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
page 34 of 93
Set the
Enable AirSaturation, Enable Temperature
, and
Enable Rawdata
to (
Yes)
to include these
parameters in the output string.
Set the
Enable Text
to (
Yes)
for the optode to output a detailed text string with the measurements, or
(
No)
to output the measured values without the descriptive text.
Set the
Enable Decimalformat
to (
Yes)
to output the measured values in decimal format. Set the
property to (
No)
to output the values in exponential format (higher resolution in the output parameter
value).
The
Interval
property describes the measurement interval in seconds; the Optode provides a set of
measured data for every measurement interval.
The
Salinity
property is set to 0 at the factory and the measurements must be post processed to
compensate for the salinity variations at the measurement site. If the salinity is known at the site and
relatively stable, you can set the Salinity property according to the known value and the measurements
will be salinity compensated inside the sensor before presentation (useful for Real-Time measurements).
The
Enable HumidityComp
property describes compensation of vapour pressure in the calculations of
the output parameters.
Enable HumidityComp
can be set to (
No)
if measurements are performed in
complete dry conditions (dry air) or if you like to perform the humidity compensation as a post-processing
operation. Measurements in dry conditions are more accurate when the
Enable HumidityComp
is set to
(
No)
. The property is set to (
Yes)
at the factory.
When
Enable Redreference
is set to (
Yes)
, the phase measurements are performed with a zero-point
set at the red reference (no fluorescence). The property can be set to (
No)
in special measurement
situations; contact AADI service department.
Enable Redreference
is set to (
Yes)
at the factory before
calibration; if the property is set to (
No)
the Optode must be recalibrated.
Set the
Enable Sleep
to (
Yes)
for the Optode to go to sleep between recordings, or (
No)
for the optode to
stay continuously switched
on
between recordings (drains more power).