Page 48
Aanderaa Data Instruments AS – TD302
7.2
Smart Sensor Terminal communication setup
Most terminal programs can be used for
Smart Sensor Terminal
communication with the sensor when connected to
a PC, e.g. Teraterm. A copy of Teraterm is included on the CD following the sensor. Copy the content of the zip file to
the Program File folder on your PC. Start the teraterm.exe file and the terminal program will start up with standard
sensor setup.
The following Smart Sensor Terminal setup should be used:
9600 Baud
8 Data bits
1 Stop bit
No Parity
Xon/Xoff Flow Control
IMPORTANT! The terminal program must send a Line Feed after each Carriage Return.
RS-232/RS-422 operation:
When connected, you will get measurement data according to the output interval stored in the flash memory. If the
measurement interval is set to zero, you can type the command
Reset
to get the operation mode and a single set of
measurement data.
7.2.1
Polled Mode:
Type a command to retrieve data/information from the sensor, refer e.g. chapter 7.1.4, chapter 7.2.6 and chapter
7.2.10for a description of available commands.
7.2.2
Sensor startup
When property
Enable Text
is set to
Yes
,
StartupInfo
is displayed at sensor power up or after reset.
StartupInfo
contains this information about product number, serial number, current mode setting, Protocol version for Smart
Sensor Terminal operation and Config Version.
Figure 7-3 Sensor Startup Information