8
The
TARE
switch will tare the scale and subtract the weight to zero as a tare weight
when the weight is a plus value. In this case the ZERO◄ and NET◄ indicators turn on.
(TARE operation) At the zero point, the net weight display shows the tare weight in
negative, and the ZERO◄ and NET◄ indicators turn on.
(
Note: In some countries or areas,
the ZERO
◄
indicator will not turn on while
the scale is tared.)
When the scale is tared, weighing range for net loads is reduced by the amount of the
tare weight.
When the ZERO operation is performed in the net weight display, the tare operation
previously done is cleared and the NET◄ indicator turns off.
(Note: In some countries or areas, the ZERO operation will not clear the TARE
operation. Press the
TARE
switch after zeroing the scale with nothing on
the weighing pan.)
Auto power-off function
If no switch is pressed and the stable indicator is displayed for a certain period of time, the
scale will automatically turn off. See the function setting
to set the elapsed time
to turn off.
When
or
is displayed (refer to
"
12.3. Error Codes
"), the auto power-off
function is enabled.
LCD backlight
The LCD backlight is controlled by the functions
and
.
If no switch is prssed and the weight display continues to be stable for a certain period of
time, the LCD backlight will automatically turn off. The elapsed time to turn off is set by the
function setting
. The backlight always on or off is also selectable.
The function setting
adjusts the brightness of the backlight.
5.3. Weight Display Resolution
The weight display resolution is a ratio of the minimum display to the weighing capacity.
The SJ-WP/-BT series has four types of weight display resolution, as shown below.
Low:
1/3,000
Normal:
1/6,000 or 1/7,500 (depending on the weighing capacity)
High:
1/12,000 or 1/15,000 (depending on the weighing capacity)
Maximum:
1/30,000
The factory setting is the normal resolution. Select the resolution according to your own
application in the function setting
.
For details about the minimum display and the weighing capacity,
refer to "
14. SPECIFICATIONS
".
The weight display resolution of the Legal for Trade models is fixed. The selection in the
function setting
is not available.
In the counting mode, the scale works with the maximum resolution regardless of the
weight display resolution selected in the function setting
.