6.Measuring Input Signals - 6.3.Starting and Ending Recording
1WMPD4004445
49
(1) Mode:
Selects the optimal mode from the nine recording modes.
(2) Data name:
Specifies the name of the recorded data.
When [Automatic numbering] is enabled, numbers are automatically appended
to the name.
(3) Recording time:
Ends recording after recording for the specified time after recording starts.
(4) Start time:
Starts recording at the specified time after starting measurement with
the START key.
(5) Interval time:
Performs recording at the specified interval. Recording ends when recording
has been performed the number of times specified in [Number of Recording
times].
(6) Recording device:
Enables/disables recording to the recording device (Printer, SSD, or Memory).
The devices can be set independently.
(7) Printer :
Enables/disables printer recording.
When enabled, the P-P data is recorded to the SSD with sampling of printer
recording.
Real-time printing:
When printer recording is enabled, real-time printing to the printer can be
enabled/disabled.
When enabled, the waveform of the specified sheet is printed from the printer
while saving the data of the printer recording.
When disabled, printing to the printer is not performed.
(8) SSD:
Enables/disables SSD recording.
(9) Memory:
Enables/disables memory recording.
Points:
Specifies the sampling count (the data count per channel) to record for each
memory recording.
Recording blocks:
Specifies the number of blocks to record for memory recording.
Endless mode:
When the endless mode is enabled, the blocks start to be overwritten from the
first block when the record blocks are full.
Pre-trigger:
Sets the pre-trigger in the memory block.
(10) Thumbnail:
Specifies the channels to display thumbnails for and the compression rate.