![background image](http://html1.mh-extra.com/html/aandd/ra3100-omniace/ra3100-omniace_simple-instruction-manual_446281051.webp)
6.Measuring Input Signals - 6.3.Starting and Ending Recording
1WMPD4004445
51
When the memory waveform is displayed on the monitor, you can display the Trigger tab from CH
on the side menu to display [Memory block] for checking the recorded memory data.
6.3.3. Pausing Recording and Scrolling Back
Start recording with the printer recorded or SSD recorded waveform displayed on the waveform monitor.
The PAUSE key can be tapped on the side menu during recording to stop the monitor waveform but
continue recording. Since the product is still in the recording state, [MEASURE] is selected on the side
menu and the waveform frame remains red. Scroll the waveform monitor to the right in this state to
scroll the waveform back and check the past waveform that has already been recorded. Scroll the
waveform monitor to the left to display the waveform recorded after pausing. You can also pinch in
(reduce)/pinch out (expand) the waveform.
Tap the PAUSE key again to monitor the waveform with the latest data.
Scroll the waveform
Pinch out (enlarge)
Pinch in (reduce)
Tap the PAUSE key to
stop the monitor.
Tap and change the display
block number with the rotary
knob.