4.Configuring Measurement - 4.2.Setup the Input Channel
30
1WMPD4004445
4.2. Setup the Input Channel
4.2.1. Channel Setup sub menu (For RA30-101)
(1) Slot number, input module type
(2) Change slot:
You can change the display slot by swiping this sub menu
left or right or tapping the left < and right > key.
(3) Select channel: Select the channel in the slot.
(4) Meas. ON/OFF
ON:
Measure and record the input signal.
(5) Color:
Change the display color of the waveform
monitor.
(6) Coupling: Switch the input signal coupling in the
order DC
→
GND
→
AC.
(7) Meas. range:
Change the measurement range of the input channel.
When this key is tapped, the rotary knob is enabled
(the LED lights up) and the range can be changed by
turning the knob.
(8) L.P.F.:
Change the low-pass filter of the input channel. When this
key is tapped, the rotary knob is enabled (the LED lights
up) and the filter can be changed by turning the knob.
(1)
(2)
(2)
(3)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
(14)
(15)
(16)
(9) A.A.F.:
Turns the anti-aliasing filter of the input channel on or off.
(10) Disp. pos.: Specify the position of the waveform monitor to display the specified range of the
waveform display area. Specified as a percentage indicating the center position of the
display range from the bottom of the monitor when the full range of the monitor is 100%.
(11) Disp. range: Specifies the display width in the amplitude direction of the waveform display area on
the waveform monitor.
Specified as the percentage of the display width when the full range of the monitor is 100%.
Example) When 50% is set, the waveform display is 10 div of the total width of 20 div.
(12) Disp. min: Set (by tapping the key and turning the knob) the display lower limit value (scale value)
of the bottom of the display range.
(13) Disp. max: Set (by tapping the key and turning the knob) the display upper limit value (scale value)
of the top of the display range.
(14) Zero adjust: Cancels the input offset of the input channel. Execute zero adjust to perform more
accurate measurement.
(15) Sheet:
Set the monitor display/printer print sheet of the set channel.
(16) The available range and current measurement value monitor.