GX-A/GF-A Maintenance manual
Page 42
Linearity and Hysteresis check
Calibrate by using the calibration mass in table below.
After calibration, in the D2 data mode, place the specified check mass in the table below one
by one on the pan, and check that the difference between the true value and the displayed
value is within the specifications (linearity). After reaching the weighing capacity, remove
each mass one by one, and check that the difference between increasing points and
decreasing points are within the specifications (
hysteresis)
.
Model
Calibration
masses
Check
masses
Linearity
(D2 mode)
Hysteresis
(D2 mode)
123 (Only GF-A)
100g
50gx2
±0.002g
±0.002g
203
200g
100gx2
303
300g
100gx3
403
400g
100gx4
603
600g
200gx3
1003
1000g
200gx5
±0.003g
±0.003g
1603
1600g
400gx4
±0.004g
1202 (Only GF-A)
1kg
500gx2
±0.02g
±0.02g
2002
2kg
1kgx2
3002
3kg
1kgx3
4002
4kg
1kgx4
6002
6kg
2kgx3
±0.03g
±0.04g
10002
10kg
2kgx5
±0.06g
6001
6kg
2kgx3
±0.1g
±0.1g
10001
10kg
2kgx5