GX-A/GF-A Maintenance manual
Page 46
6.11 Parameter Settings
1. Serial number setting
Follow the procedure below to set the serial number.
Step 1.
Display the function setting mode
Unit ,
after getting into check mode shown
on page 28,29.
Step 2. Press the SAMPLE key twice.
5n
is displayed.
Step 3. Press the PRINT key. Then the display starts to blink.
Step 4. Using the following keys, input the serial number on the label attached to the each
balances.
RE-ZERO key: changes the value of the blinking digit
(+)
.
MODE key:
changes the value of the blinking digit
(-)
.
SAMPLE key: changes the blinking digit position.
CAL key :
cancels the operation
Step 5. When the serial number is set, press the PRINT key.
Step 6.
end
is displayed to indicate that the operation is completed.
Summary of Contents for GF-A Series