62
Using the "
UN:mm
" command, the unit mass can be recalled.
The unit mass recalled can be output using the "
?UW
" command.
The unit mass can be changed using the "
UW
:" command.
"
mm
" indicates a two-digit numerical value 01 to 50, which corresponds to
p01
-
p50
of the
GX-K series.
"
mm
" indicates a two-digit numerical value 01 to 20, which corresponds to
p01
-
p20
of the
GF-K series.
Weighing input mode
In the weighing input mode, the specified number of samples is placed on the pan to store
the unit mass. Re-storing the unit mass or performing Automatic Counting Accuracy
Improvement (ACAI) on the re-stored unit mass is possible. Follow the procedure described
in "6-2. Counting Mode (PCS)".
Use the following keys to store a unit mass in the weighing input mode.
RE-ZERO
key.. To set the display to zero.
10
-
pcs
→
10
0
pcs
SAMPLE
key ... To change the number of samples to be stored.
10
0
pcs
→
25
0
pcs
key....... Press, after a sample is placed, to store the unit mass. Go to step 3 of the
previous page.
CAL
key.......... To return to the unit mass confirmation mode. Go to step 3 of the
previous page.
MODE
key....... Press and hold to go to the digital input mode.
Digital input mode
To use this mode, the sample unit mass must be known
beforehand. In the digital input mode, the unit mass value is
entered digitally using the keys. The display in the digital input
mode is shown to the right.
Note ACAI can not be used on the unit mass stored
using the digital input mode.
Use the following keys to store a unit mass in digital input mode.
SAMPLE
key ... To select the digit to be changed.
RE-ZERO
key.. To change the value of the selected digit.
MODE
key ...... To change the decimal point position.
key....... To store the unit mass. Go to step 3 of the previous page.
CAL
key.......... To return to the unit mass confirmation mode. Go to step 3 of the
previous page.
MODE
key....... Press and hold to go to the weighing input mode.
Note If the new unit mass is out of the setting range, "
error 2
" is displayed.
Refer to "20. Specifications" for the minimum unit mass.
Value
Digit
Decimal point
Store