22
7. Weighing Speed Adjustment / Self Check Function
7-1. Weighing Speed Adjustment
This function detects the influence on weighing that is caused
by drafts and/or vibration at the place where the balance is
installed and sets the following three weighing speeds
(response characteristics) automatically.
The function has three rates as follows:
Changing the weighing speed changes the display refresh rate.
Indicator Parameter Weighing
Speed Stability
Display refresh rate
FAST
Cond
0
MID.
Cond
1
SLOW
Cond
2
Fast response, Sensitive value
Slow response,
Stable value
If the weighing speed is changed as follows:
MID. or SLOW FAST =10 times/second
FAST MID. or SLOW = 5 times/second
Step 1 Press and hold the
MODE
key until
RESPONSE
is
displayed. And then, press the
MODE
key again
quickly.
Step 2 Press the
MODE
key to select a weighing speed.
Either
FAST
,
MID.
or
SLOW
can be selected.
Step 3 After a few seconds of inactivity the balance displays
end
.
Then, it returns to the weighing mode and displays the
updated response indicator. The response indicator
remains displayed for a while.
Note
The weighing speed adjustment can be changed at "Condition (
Cond
)" of
"Environment, Display (
ba5fnc
)" in the function table. Refer to "10. Function
Table" for details.
To set a refresh rate of 5 times/second when the response rate is
FAST
or
10 times/second when the response rate is
MID.
or
SLOW
, change the
"Display refresh rate (
5pd
)" parameter of "Environment, Display (
ba5fnc
)" in
the function table.
Weighing speeds
Release and wait
Press again
Press and hold
Release
Select a parameter
with pressing it.
Summary of Contents for GF-10K
Page 7: ...5 ...
Page 95: ...93 20 1 External Dimensions Unit mm ...
Page 102: ...100 MEMO ...
Page 103: ......