86
18. Maintenance
18-1. Treatment Of The Balance
In normal use, the balance can be cleaned with water. But, keep the following
precautions so that dust and water do not invade the balance.
Do not direct water pressure at the bottom of the balance.
Do not use powerful water jets.
Do not submerge the balance in water.
Clean the balance with a lint free cloth that is moistened with warm water and a mild
detergent.
Do not use organic solvents to clean the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs
service or repair.
Use the original packing material for transportation.
While cleaning the balance and keeping it waterproof, connect a waterproof RS-232C
cable (GX-07K) or cover terminals of the RS-232C interface and AC adapter jack. Insure
that the underhook cover is in place.
The value of the internal mass may change due to aging, corrosion or other damage
caused by the operating environment. Check the internal mass periodically. Correct the
internal mass value as necessary.
Weighing pan
Pan support
Draft gate
RS-232C interface
Cover the terminal or connect the waterproof RS-232C
cable (GX-07K) to keep waterproof and dustproof.
AC adapter jack
When the AC adaptor is not connected, cover the terminal.
Handling hole
Summary of Contents for GF-10K
Page 7: ...5 ...
Page 95: ...93 20 1 External Dimensions Unit mm ...
Page 102: ...100 MEMO ...
Page 103: ......
Page 104: ......