Balance Software Version 1.660 or later
Three settings are available in “Password function
(
LocFnc
) ” of the function table.
pA55
0
No password function.
pA55
1
Entering a password is required before starting weighing.
pA55
2
Logging in by the administrator password is required before making changes.
pA55
0
(No password function)
No password is required. Anybody can perform weighing, use all the balance functions and make
changes to them.
pA55
1
(Entering a password is required before starting weighing.)
The administrator (
ADM
IN
) sets unique passwords to restrict balance users. Pressing the
ON:OFF
key before weighing will prompt you to enter a password.
The correct password must be entered for the balance to enter the weighing mode.
Two login levels are available: administrator (
ADM
IN
) and user (
U5ER
01
to
10
).
All the functions and settings can be used.
Administrator (
ADM
IN
)
A password can be set for each of 10 users.
User (
U5ER
01
to
10
)
Initialization and changes to the settings are restricted,
including the clock setting.
No password
The balance cannot be used.
pA55
2
(Logging in by the administrator password is required before making changes.)
Anybody can perform weighing, but initialization and changes to the settings are restricted,
including the clock setting. Even if the
ON:OFF
key is pressed to start weighing, a password
will not be required.
Two login levels are available: administrator (
ADM
IN
) and guest (
GUE
5t
).
Administrator (
ADM
IN
)
All the functions and settings can be used.
Guest (
GUE
5t
)
*No password
Initialization and changes to the settings are restricted,
including the clock setting.
When weighing is started by pressing the
ON:OFF
key while holding down the
CAL
key in
the standby mode, the balance will prompt you to enter the administrator password.
Items restricted depending on the login level
Weighing
Login level
Entering password
before weighing
Calibration
Changes to the settings
*1
Administrator
(
ADM
IN
)
Available Available
User
(
U5ER
01
to
10
)
Required
Guest
(
GUE
5t
)
Not required
Available or
not available
*2
Not available
*1
: Response adjustment, repeatability test, function switch and initialization, function settings
(clock setting)
*2
: Available under ordinary circumstances, but not available to users and guests when the
administrator sets the function switch to “inhibit”.