11.2.
GLP Report
To output the GLP/GMP compliant report to an AD-8121B printer, AD-8126 printer, AD- 8127
printer or to a personal computer, set the "GLP output (
info
)" parameter of the function table to
"
1
" (AD-8121 format), "
2
" (General format using the balance built-in clock data) or "
3
" (General
format using the external device clock data).
Notes on outputting the GLP/GMP compliant report to a printer
Refer to "
16.2. Connecting to Peripheral Equipment
" for connection to a printer.
With an AD-8121B printer, use MODE 3. If MODE1 is used, select temporarily the dump print mode
by pressing the
STAT.
key of the AD-8121B printer.
With an AD-8127 printer, use the dump printing mode.
If the external key printing mode is used, press and hold the
ENT
key of the AD-8127 printer (for
approx. 2 seconds) to switch between the external key printing mode and the dump printing mode.
If the time and date are not correct when the balance built-in clock data is output (
info 1
,
2
), set the
correct time and date in "Clock (
Cl
adj
)" of the function table.
“
info 3
”
is available for balances with software version 1.660 or later.
Calibration report using the internal mass
Key operation
1 Press
the
CAL
key to display
Cal
in
. The balance performs calibration using the internal mass.
2 If GLP output is used,
glp
is displayed and the calibration report is output.
3 The balance returns to the weighing mode.
Function
setting
"
info
1
" Function
setting
"
info
2
"
AD-8121
format
General format using the balance built-in clock data
Space,
ASCII
20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Signature
===========
A
=
&
=
D
MODEL
=====
BM-300
S/N
=====
01234567
ID
======
LAB-0123
DATE
==
2011/02/20
TIME
====
12:34:56
CALIBRATED(INT.)
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-