61
Calibration test report using the internal mass
Note
Calibration test does not perform adjustment.
Key operation
1 Press and hold the
CAL
key (for approx. 2 seconds) to display
CC
in
and release the key.
2
CC
is displayed and the balance is checked automatically.
3 The zero point is measured and the weight value is displayed for a few seconds.
4 The internal mass is weighed and the weight value is displayed for a few seconds.
5
glp
is displayed and the calibration test report is output.
6 The balance returns to the weighing mode.
Command
This calibration test can be performed with the TST command.
Function
setting
"
info
1
" Function
setting
"
info
2
"
AD-8121
format
General format using the balance built-in clock data
Space,
ASCII
20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test type
Zero point value
Target weight value
Target weight
Signature
-----------
A
-
&
-
D
MODEL
-----
BM-300
S/N
-----
01234567
ID
------
LAB-0123
DATE
--
2011/02/20
TIME
----
12:34:56
CAL.TEST(INT.)
ACTUAL
-------
0.0000
--
g
----
+200.0002
--
g
TARGET
----
+200.0000
--
g
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-