21
9-5
Detailed information of the Input data2
(
Only for AD-4212C
/
AD-
4212D
)
The following items change only when connected to the AD-4212C / AD-4212D.
When connected to other weighing devices, it will be the initial value.
■
Response characteristics [CH1
~
4 Response characteristics] (CH1:byte 10,11, CH2:byte 28,29, CH3:46,47,
CH4:byte 64,65)
Value
Status of the response
characteristics
※1
0x0000
Initial value
0x0001
FAST
0x0002
MID
0x0003
SLOW
0x0004
User setting
※2
※1.
You can switch between FAST / MID / SLOW with the "U" command.
(
Refer to “9-2. Output data”
)
※2.
With AD-4212C, the response characteristics are set in detail.
(Refer to the WinCT-AD4212C additional instruction manual for the setting method)
■
Adjustment status [CH1 to 4 Adjustment status] (CH1:byte 12,13, CH2:byte 30,31, CH3:byte 48,49, CH4:byte 66,67)
Value
Status
0x0000
Initial value
0x0001
Waiting for zero point weighing
0x0002
Weighing the zero point
0x0003
Waiting for adjustment weight to be
weighed
0x0004
Weighing adjustment weight
0x0005
Sensitivity adjustment completed
0x0006
Sensitivity adjustment error
■
Adjustment weight value [CH1 to 4 Adjustment weight value]
(CH1:byte 14 to 17, CH2:byte 32 to 35, CH3:byte 50 to 53, CH4:byte 68 to 71)
You can check the value of the calibration weight to be used when performing calibration (sensitivity adjustment) with
your own weight.
‐
The unit is grams and the decimal point is a value without a decimal point.
‐
The value of the calibration weight to be used is used only when "Waiting for calibration weight to be weighed".
The initial value is "0xFFFFFFFF". When you return to the weighing mode, it will return to the initial value.
Example of Adjustment weighing value for weighing device (CH1)
Weight value
Byte
Name
Contents
Data
Remarks
100g
14 to 17
CH1 Adjustment weight value
Adjustment
weight
value
0x00000064
⇒
100