![background image](http://html1.mh-extra.com/html/aandd/ad-8552eip/ad-8552eip_instruction-manual_3074539020.webp)
20
9-4
Detailed information of the Input data
1
■
Weighing value [CH1 to 4 Weighing value] (CH1:byte 0 to 3, CH2:byte 18 to 21, CH3:byte 36 to 39
、
CH4:byte 54 to 57),
Decimal point position [CH1 to 4 Decimal point position] (CH1: byte 6,7 CH2: byte 24,25, CH3: byte 42,43, CH4: byte 60,61)
Example of weighing value and decimal point value by output of Weighing device (CH1)
(
When the decimal point
position setting by SW-5 is AUTO
)
Weighing value
of the weighing
device (CH1)
Byte
Name
Contents
Data
Remarks
123.456g
0 to 3 CH1 Weighing value
Weighing value 0x0001E240
⇒
123456
6,7
CH1 Decimal point position Decimal point
position
0x0003
⇒
3
123.456g
0 to 3 CH1 Weighing value
Weighing value 0x00003039
⇒
12345
6,7
CH1 Decimal point position Decimal point
position
0x0002
⇒
2
■
Status information of the weighing device [CH1 to 4 Status information] (CH1:byte 4,5, CH2:byte 22,23, CH3:byte 40,41,
CH4:byte 58,59)
Byte
Status
Remarks
0
Stable / unstable status of weighing
value
0
:
Unstable
1
:
Stable
1
Weighing range over flag
1
:
When the weighing value exceeds the maximum display
2
Non-weighing status flag
0
:
When receiving the current weighing value from the weighing device
1
:
When weighing value is interrupted for 2 seconds or more, and rezero
or calibration is in progress.
3
Rezeroing flag
1
:
Rezero processing in progress
※1
4
No function
5
Adjustment in progress flag
1
:
Adjustment in progress
※1
6
Adjustment error flag
※2
1
:
When calibrating If the value of the weighing calibration weight is not
appropriate.
(
This flag returns to 0 when shifting to the weighing mode.
)
7
No function
8 to 15
※
1
.
The flag does not change when operated with the keys of the weighing device.
※
2
.
This flag can be used only when connected to the AD-4212C / D.
Caution:
Be sure to read "Weighing value" and "Weighing value status information" at the same time.
If they
are read separately, the correspondence between the "weighing value" and the "weighing value
status information" may differ depending on the update timing, so the weighing value status cannot
be determined accurately.
■
Weighing unit [CH1
~
4 Weighing unit] (CH1:byte 8,9, CH2:byte 26,27, CH3:byte 44,45, CH4:byte 62,63)
Relationship between value and unit
Value
Unit (mode)
Value
Unit (mode)
Value
Unit (mode)
0x0000 g (gram)
0x0006
Mom (Momme)
0x000C TL (Tael)
0x0001 mg (milligram)
0x0007
OZ (Ounce)
0x000D Tol (Tola)
0x0002 kg (kilogram)
0x0008
Lb (Pound)
0x000E
MES (Messghal)
0x0003 PCS (pieces)
0x0009
OZt (Troy Ounce)
0x000F
N (Newton)
0x0004 % (percent)
0x000A dwt (pennyuweight)
0x0010
MLT (Programmable unit)
0x0005 ct (carat)
0x000B GN (Grain)
0xFFFF Indistinguishable