19
9-3
Input data (Instance100)
The data transferred from AD-8552EIP to PLC, etc. is called Input data.
Input data is used when acquiring data from the weighing device.
■
Input data layout (T->O)
[ Weighing device (CH1 to 4) -> AD8552EIP -> PLC, etc. ]
Assembly
Instance Attribute
Byte
Name
Data type
Contents
100
3
0
~
3
CH1 Weighing value
Int32
Weighing value
Weighing
device
CH1
4,5
CH1 Status information
UInt16
Status information
※
1
6,7
CH1 Decimal point position
UInt16
Decimal point
position
8,9
CH1 Weighing unit
UInt16
Weighing unit
10,11 CH1 Response characteristics UInt16
Response
characteristics
12,13 CH1 Adjustment status
UInt16
Adjustment status
14
~
17 CH1 Adjustment weight value
Int32
Adjustment weight
value
※
2
18
~
21 CH2 Weighing value
Int32
Weighing value
Weighing
device
CH2
22,23 CH2 Status information
UInt16
Status information
※
1
24,25 CH2 Decimal point position
UInt16
Decimal point
position
26,27 CH2 Weighing unit
UInt16
Weighing unit
28,29 CH2 Response characteristics UInt16
Response
characteristics
30,31 CH2 Adjustment status
UInt16
Adjustment status
32
~
35 CH2 Adjustment weight value
Int32
Adjustment weight
value
※
2
36
~
39 CH3 Weighing value
Int32
Weighing value
Weighing
device
CH3
40,41 CH3 Status information
UInt16
Status information
※
1
42,43 CH3 Decimal point position
UInt16
Decimal point
position
44,45 CH3 Weighing unit
UInt16
weighing unit
46,47 CH3 Response characteristics UInt16
Response
characteristics
48,49 CH3 Adjustment status
UInt16
Adjustment status
50
~
53 CH3 Adjustment weight value
Int32
Adjustment weight
value
※
2
54
~
57 CH4 Weighing value
Int32
Weighing value
Weighing
device
CH4
58,59 CH4 Status information
UInt16
Status information
※
1
60,61 CH4 Decimal point position
UInt16
Decimal point
position
62,63 CH4 Weighing unit
UInt16
Weighing unit
64,65 CH4 Response characteristics UInt16
Response
characteristics
66,67 CH4 Adjustment status
UInt16
Adjustment status
68
~
71 CH4 Adjustment weight value
Int32
Adjustment weight
value
※
2
※
1
.
If you connect to anything other than the AD-4212C / D, the functions will be limited.
※
2
.
This function can be used only when connected to the AD-4212C / D.