Example of calibration (sensitivity adjustment) with external weight
(
Possible only with AD-4212C
/
AD-4212D
)
Steps
Items
Contents
Input data layout
(Instance150 ,Attribute3)
Output data layout
(Instance100 ,Attribute3)
Status confirmation contents
AD-4212C
AD-4212D
1
Instructs to start the sensitivity
adjustment mode
Sets "1" to the corresponding data layout
Byte 0,1 [CH1 Command]
bit 1
‟
CAL”
bit 0
‟
EXC”
2
Confirms the sensitivity
adjustment mode
Refer to the corresponding data layout.
Byte12,13 [CH1 Adjustment status]
Wait until 0x0001
※1
3
Instructs to start weighing at zero
point
Sets "1" to the corresponding data layout Byte 0,1 [CH1 Command] bit 2
‟
PRT”
4
Confirms to start weighing at
zero point
Refer to the corresponding data layout.
Byte12,13 [CH1 Adjustment status]
Wait until 0x0002
※1
5
Confirms during weighing at the
zero point
Refer to the corresponding data layout.
Byte12,13 [CH1 Adjustment status]
Wait until 0x0003
※2
6
Command clear
Sets "0" to the corresponding data forma Byte0,1 [CH1 Command] bit 2
‟
PRT”
7
Confirms the value of the sensitivity
weight to be used
※4
Refer to the corresponding data layout.
Byte14-17
[CH1 Adjustment weight value]
Weight value to use
(
eg
)
0x000000C8
⇒
200g
8
Places the weight
9
Instructs to start weighing the
weight
Sets "1" to the corresponding data layout Byte0,1 [CH1 Command] bit 2
「
PRT
」
10
Confirms to start weighing weight Refer to the corresponding data layout.
Byte12,13 [CH1 Adjustment status]
Wait until 0x0004
※1
11
Confirms during weighing weight
Refer to the corresponding data layout.
Byte12,13 [CH1 Adjustment status]
Wait until 0x0005
※2※3
12
Replaces the weight
13
Confirms the transition to the
weighing state
Refer to the corresponding data layout.
Byte 0-3 [CH1 Weighing value]
Wait until 0x00000000
14
Command clear
Sets "0" to the corresponding data layout
0,1 [CH1 Command] bit 2
‟
PRT”
bit 1
‟
CAL”
bit 0
‟
EXC”
※
1
.
Normally, it changes immediately after setting "1" to the bit of the previous step.
※
2
.
If the weighing device is unstable, the sensitivity adjustment will not be executed and the weighing state will automatically shift to the weighing state (0x00000000).
※
3
.
If the value of the weighed sensitivity adjustment weight is not appropriate, a sensitivity adjustment error (0x0006) will occur. After that, it automatically shifts to the weighing
state (0x00000000).
※
4
.
If you want to change the value of the sensitivity adjustment weight to be used, please check the instruction manual of the relevant weighing device.
■
Calibration (sensitivity adjustment) execution example using the built-in weight
(
Only with the AD-4212D
)
Steps
Items
Contents
Input data layout
(Instance150 ,Attribute3)
Output data layout
(Instance100 ,Attribute3)
Status confirmation contents
AD-4212D
1
Calibration (sensitivity
adjustment) execution
example using the built-in
weight
Sets "1" to the corresponding data layout Byte0,1 [CH1 Command] bit 1
‟
CAL”
2
Confirms to start the sensitivity
adjustment
Refer to the corresponding data layout.
Byte 4,5 [CH1 Status information]
Wait until bit 5 becomes 1
※1
3
Confirms the completion
Refer to the corresponding data layout.
Byte 4,5 [CH1 Status information]
Wait until bit 5 becomes 0
※1
Byte 12,13 [CH1 Adjustment status]
Confirms becomes 0x0005
※5
4
Command clear
Sets "0" to the corresponding data layout Byte0,1 [CH1 Command] bit 1
‟
CAL”