AWG5200 Series

Arbitrary Waveform Generators

Declassi

fi

cation and Security

ZZZ

Instructions

xx

www.tektronix.com

*P077133800*

077-1338-00

Summary of Contents for AWG5202

Page 1: ...AWG5200 Series Arbitrary Waveform Generators Declassification and Security ZZZ Instructions xx www tektronix com P077133800 077 1338 00...

Page 2: ...ued and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of...

Page 3: ...Table of Contents Preface iii Clear and Sanitize Procedures 1 Memory devices 1 Media storage and data export devices 5 AWG5200 Series Declassification and Security Instructions i...

Page 4: ...Table of Contents ii AWG5200 Series Declassification and Security Instructions...

Page 5: ...ocument are written to meet the requirements specified in NISPOM DoD 5220 22 M Chapter 8 ISFO Process Manual for Certification Accreditation of Classified Systems under NISPOM Products The following T...

Page 6: ...e instrument For the purpose of clearing data putting the instrument in Standby mode does not qualify as powering off For these products you will need to either press a rear panel OFF switch or remove...

Page 7: ...User settings Describes the type of information stored in the device Refers to instrument settings that can be changed by the user Both Describes the type of information stored in the device It means...

Page 8: ...ations System board No Remove the power source from the instrument for at least 20 seconds Table 2 Nonvolatile memory devices Type and min size Function Type of user info stored Method of modification...

Page 9: ...lds all user storable data such as waveforms measurement results and instrument settings User data user settings Indirect Firmware operations user input Front panel Yes To clear see footnote 2 To sani...

Page 10: ...tion Use this procedure to restore the operating system and the AWG application on your instrument to the original factory condition This procedure does not erase or change factory calibration constan...

Page 11: ...g table lists the media storage and data export devices in the instrument and listed options Detailed procedures to disable these devices if any are shown following the table Table 3 Media and data ex...