![background image](http://html1.mh-extra.com/html/aandd/ad-4212a/ad-4212a_instruction-manual_446302042.webp)
40
L A B - 0 1 2 3
CR LF
2 0 0 4 / 0 7 / 0 1
CR LF
CR LF
1 2 : 3 4 : 5 6
LAB-0123, No,012, 2004/07/01, 12:34:56, ST,+000.1278,
~~
g<CR><LF>
CR LF
S T , + 0 0 0 . 1 2 7 8 ,
g
CR LF
O L , + 9 9 9 9 9 9 9 E + 1 9 ,
g
CSV format
5if type 5
This format separates the data of A&D standard format and the unit by a comma ( , ).
This format outputs the unit even when the data is overloaded.
When a comma ( , ) is selected for decimal point, separators are set to semicolon ( ; ).
When the ID number, data number, time and date are added at "Data output (
dout
)" of the
function table, outputs ID number, data number, date, time and weighing data in this order and
separates each item by a comma and treats all the items as one group of data.
10-7 Description of the Data Format Added to the Weighing Data
ID number
dout 5-id 1
The number to identify a specific balance.
This format consists of eight characters excluding the terminator.
Data number
dout d-no 1
This format outputs the data number just before the data is transmitted using the RS-232C interface.
This format consists of six characters excluding the terminator.
When CSV format (
5if type 5
) is selected, the period ( . ) is replaced with a comma ( , ).
Date
dout 5-td 2 or 3
The date output order can be changed in " Time/Date output (
5-td
)" and "Clock (
Cl adj
)".
The year is output in a four-digit format.
Time
dout 5-td 1 or 3
This format outputs time in 24-hour format.
Data number
Terminator
ID number
Data number
Date
Time
Weighing data