![A&D AD-4212A Instruction Manual Download Page 86](http://html1.mh-extra.com/html/aandd/ad-4212a/ad-4212a_instruction-manual_446302086.webp)
84
23. SPECIFICATIONS
AD-4212A-100 AD-4212A-200 AD-4212A-600
AD-4212A-1000
Weighing capacity
110 g
210 g
610 g
1100 g
Maximum display
110.0084 g
210.084 g
610.084 g
1100.084 g
Minimum weighing value (1 digit)
0.1 mg
1 mg
Repeatability (Standard deviation)
0.15 mg
1 mg
Linearity
±
0.3 mg
±
2 mg
±
3 mg
Stabilization time in seconds
(typical at FAST under good
environment)
0-30 g
1.1
0-30 g 0.8
0-30 g
0.9
0-30 g
0.9
30-110 g 1.3 30-210 g 1.0 30-610 g 1.1
30-1100 g
1.1
Display refresh rate
10 times/second (5 times/second can be selected)
I/O unit RS-232C
Bi-directional, 600-19200 bps
Comparator contact
output
3-level output
External contact input
RE-ZERO, PRINT
Buzzer
Sounds the connected buzzer for 3-level comparison results
(It becomes 5-level when OP-04 is installed)
Sensitivity drift
±2 ppm/°C (10°C-30°C/50
F-86
F )
Operating environment
5°C to 40°C (41°F to 104°F), 85% RH or less (No condensation)
Calibration weight provided
(Conforming to OIML Class E2)
50 g
100 g
200 g
Applicable weight values
20 g, 50 g, 100 g
20 g, 50 g,
100 g, 200 g
20 g, 50 g, 100 g,
200 g, 300 g, 400 g,
500 g, 600 g
20 g, 50 g, 100 g, 200 g,
300 g, 400 g, 500 g, 600 g,
700 g, 800 g, 900 g, 1000 g
Weighing
unit
Dimensions
80 (W)
×
230 (D)
×
90 (H) mm
80 (W)
×
230 (D)
×
92.5 (H) mm
Weighing pan
50
×
50 mm
70
×
70 mm
Mass
Approx. 2 kg
Display Dimensions
237 (W)
×
150 (D)
×
155 (H) mm
Mass
Approx. 2 kg
Connection cable
Approx. 2 m (between weighing unit and display)
AC adapter
Confirm that the adapter type is correct for the local voltage and power receptacle type
Power consumption
Approx. 15VA (supplied to the AC adapter )
Counting
mode
Minimum unit mass
0.1 mg
1 mg
Number of samples
10, 25, 50 or 100 pieces
Percent
mode
Minimum 100%
reference mass
10.0 mg
100 mg
Minimum 100%
display
0.01%
、
0.1%
、
1%
(Depends on the reference mass stored.)
Options
OP-01
BCD output, external contact input
(
RS-232C and comparator contact output disabled
)
OP-04
RS-232C, 5-level comparator contact output, external contact input
(Not field installable. Must be ordered with the balance for installation at the factory.)
OP-07
*
Extension cable 3 m
OP-08
Ethernet interface
(
RS-232C and comparator contact output disabled
)
OP-19
Stainless steel breeze break
(
Provided as standard for AD-4212A-100
)
OP-20
Metal leveling foot
Only one extension cable can be used. The maximum distance between the weighing unit
and the display unit is 5 m (connection cable 2 m + extension cable 3 m).
Summary of Contents for AD-4212A
Page 8: ...6...
Page 9: ...7...
Page 79: ...77...
Page 104: ...102 Z Zero point 25 27 Zero point drift 36 Zero tracking 31 36...
Page 105: ...103 MEMO...
Page 106: ...104 MEMO...
Page 107: ......